Application: Camsafe chambers are used to separate radioactive, toxic or bacterial particles and gases for maximum operator safety.
Areas of use include laboratories, isolation wards and the pharmaceutical industry. The casings can be joined together to achieve multi-stage filter systems with a volume flow of up to 24,000m2/h.
Epoxy coated casing: Epoxy resin coating in RAL 9010 which can be decontaminated, clamping device in high grade steel 1.4301.
Steel casing: Manufactured in high grade steel 1.4301.
Standard design: Material thickness 2 mm, flange drilled ready for operation, service cover with male thread hand knobs, changing bag, support shelf for contact-free filter changing, automatic tension-regulating filter clamping device.
Options: Filter seal seating test device in accordance with DIN 1946 Part 4, pressure compensation device and pressure check points
Filter cartridges:
Prefilter P1: 610 x 610 x 47 mm
Prefilter P2: 610 x 610 x 292 mm
Main filter: 610 x 610 x 292 mm
All main filters come equipped with straps to facilitate handling and changing.
The filters can be supplied with a variety of gaskets to suit the application.
The casing is suitable for various classes of filter, and for dealing with gases or
particle contamination.
Accessories: Safe change bags, manometers and material for the flanged joints
supplied on request.
Benefits of Camsafe:
- Contact-free filter change
- Automatic tension-regulating filter clamping device
- Fast, secure filter installation, lever-operated
- Optional filter seal seating test as per DIN 1946, Part 4
- Sealed, welded, solid design
- Tightness class B as per EN 1866 at 5,000 Pa
- Ready-flanged
- CamSafe connecting pieces can be supplied in various designs